Journal of the American College of Nutrition, Vol 15, Issue 4 397-402, Copyright © 1996 by American College of Nutrition
JOURNAL ARTICLE |
D. C. Frankenfield, G. Y. Sarson, S. A. Blosser, R. N. Cooney and J. S. Smith
Department of Clinical Nutrition, Pennsylvania State University, Milton S. Hershey Medical Center, Hershey 17033, USA.
OBJECTIVE: Numerous protocols are used for indirect calorimetry in research and clinical settings. The objective of the current study was to validate in critically ill patients an abbreviated protocol that uses five consecutive stable 1-minute readings of oxygen consumption (VO2), carbon dioxide production (VCO2), and minute ventilation (VE) in a range of +/- 5%, versus a more standard protocol that uses 30 consecutive stable one minute readings of VO2, VCO2, and VE in a range of +/- 10%. METHODS: Indirect calorimetry was performed on resting, mechanically ventilated, critically ill patients. The first 5-minute period in which coefficients of variation for VO2, VCO2, and VE were < or = 5% was compared to the first 30-minute period in which coefficients of variation for these variables were < or = 10%. RESULTS: Thirty-four critically ill patients were studied. Twenty four patients (70%) successfully completed both protocols (Success Group). Eighteen percent of subjects completed neither the abbreviated nor the 30 minute protocol, and 12% completed only one protocol (Fail Group). The Success Group was marked by a higher incidence of sedation and/or medical paralysis. There were no significant differences in VE, VO2, VCO2, respiratory quotient, or energy expenditure between the protocols in the Success Group or the Fail Group. Coefficients of determination (R2) for VO2 and VCO2 between the two methods in the Success Group were 0.99. In the Fail Group, R2 values ranged from 0.75 for VCO2 to 0.91 for VO2. CONCLUSION: In sedated, mechanically ventilated patients, an indirect calorimetry test of five consecutive 1-minute periods with coefficient of variation < or = 5% is equivalent to a longer test consisting of 30 consecutive 1-minute periods with coefficient of variation < or = 10%.
This article has been cited by other articles:
![]() |
C. McEvoy, S. R. Cooke, and I. S. Young A Reduced Abbreviated Indirect Calorimetry Protocol Is Clinically Acceptable for Use in Spontaneously Breathing Patients With Traumatic Brain Injury Nutr Clin Pract, August 1, 2009; 24(4): 513 - 519. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. C. Frankenfield, A. Coleman, S. Alam, and R. N. Cooney Analysis of Estimation Methods for Resting Metabolic Rate in Critically Ill Adults JPEN J Parenter Enteral Nutr, January 1, 2009; 33(1): 27 - 36. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. D. Branson and J. A. Johannigman The Measurement of Energy Expenditure Nutr Clin Pract, December 1, 2004; 19(6): 622 - 636. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S. White, R. W. Shepherd, and J. A. McEniery Energy Expenditure Measurements in Ventilated Critically III Children: Within- and Between-Day Variability JPEN J Parenter Enteral Nutr, September 1, 1999; 23(5): 300 - 304. [Abstract] [PDF] |
||||
![]() |
D. C Frankenfield, R. N Cooney, J S. Smith, and W. A Rowe Bioelectrical impedance plethysmographic analysis of body composition in critically injured and healthy subjects Am. J. Clinical Nutrition, March 1, 1999; 69(3): 426 - 431. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. C. Frankenfield, J S. Smith, and R. N. Cooney Accelerated Nitrogen Loss After Traumatic Injury Is Not Attenuated by Achievement of Energy Balance JPEN J Parenter Enteral Nutr, November 1, 1997; 21(6): 324 - 329. [Abstract] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |